ReLeaf-logo

ReLeaf RESTORATIVE Basic Setup Kit

ReLeaf-RESTORATIVE-Basic-Setup-Kit-product

Texnik xususiyatlari

  • Material: Soft, flexible thermoplastic elastomer
  • 100% BPA and latex-free
  • Intended Use: Restorative dental procedures

MAHSULOT TAVSIFI/FOYDALANIShI MAQSADASI

ReLeaf® Restorative is a hands-free, high-volume dental suction device that assists practitioners with evacuation, isolation, retraction, and maintaining a dry field during general dentistry, restorative, and hygiene procedures. ReLeaf® is for clinical use by dental professionals.
The Leaf mouthpiece comprises a soft, flexible thermoplastic elastomer that is 100% BPA and latex-free. No health hazards have been identified with the use of this material.
The ReLeaf® hoses comprise of a silicone rubber (silicone and silicon dioxide).

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (1)

O'RNATISH VA KEMALLARNING FOYDALANISHI

  1. Fold the two Leaf halves together by pushing together the small protruding knob into the opposing hole at the Restorative Leaf base. Insert the larger knob into the opposing hole at the Leaf tip. Squeeze the two halves of the Leaf together to ensure each side securely attaches to the other. Attach the Leaf to the Restorative Adapter on the Short Hose (8” Quick Disconnect Hose) with the smaller leaf face facing away from the hose. Ensure the Leaf is securely attached.
  2. Insert the ReLeaf® Bite Block onto the peg closest to the Leaf on the Restorative Adapter. Making sure the concaved slope is in contact with the Restorative Leaf.ReLeaf-RESTORATIVE-Basic-Setup-Kit- (2)
  3. Attach the free end of the
    Short Hose to the Quick Disconnect Adapter on the Long Hose (48” HVE Attachment Hose). Insert the coupler at the other end of the Long Hose securely into an open HVE port. It may be necessary to wet the
    O- ring inside the unused port so the Adapter can slide in easily.
  4. Place the ReLeaf® Bite Block on the opposing arch you are working on. While having the Leaf sweep in front of the tongue and wrap around the retromolar area to the opposing cheek side.
  5. When transitioning between patients, a cleaned and sterilized Bite Block, Restorative Adapter, and Short Hose must be utilized before attaching a new Leaf.ReLeaf-RESTORATIVE-Basic-Setup-Kit- (3)

QO'SHILMALARI
Ma'lum kontrendikatsiyalar yo'q.

EHTIYOT CHORALARI

Risk of allergic reactions cannot be excluded for those predisposed to allergies.

STERILIZATION, DISPOSAL AND STORAGE

All gray and white components of the ReLeaf™ system are compatible with standard dental sterilization protocols. These parts can be: wiped with dental-approved disinfectant wipes or sprays, run through dental atomizer, processed through an ultrasonic bath, run through a thermal washer, and run through autoclave cycle per sterilization cycle table.

  • Barglari: bir martalik, bir martalik
    Bir martali ishlatiladigan barglar faqat bir martalik foydalanish uchun mo'ljallangan. Barglarning materiali va dizaynining tabiati tufayli ularni sterilizatsiya qilish va qayta ishlatish kerak emas. Yuqumli patogenlar tarqalishining oldini olish uchun barglarning og'iz bo'shlig'i klinik chiqindi idishiga to'g'ri tashlanishi kerak.
  • Bite Block, Restorative Adapter and Short Hose
    100% Autoclavable and Reusable.
    Before placement in an ultrasonic cleaner, rinse the Bite Block, Restorative Adapter and Short Hose with water and brush both external and lumen surfaces with a soft bristle brush for the exterior surface and a soft bristle lumen brush or pipe cleaner for the interior surfaces. Place in an ultrasonic unit filled with enzymatic cleaner. (note: follow instructions with enzymatic cleaner for proper mixing, exposure time, etc., per manufacturer instructions). Rinse with water, place in an autoclavable pouch. Run product through a wrapped / pouched cycle in the autoclave.
  • Long Hose, Quick Disconnect Adapter and HVE Adapter
    100% Autoclavable and Reusable.
    The Long Hose should stay connected to the HVE until ready for routine evacuation cleaning with an enzymatic evacuation cleaner. Running evacuation cleaners through the hose to clean out any debris at least once per week is recommended. It is not necessary to run evacuation cleaners through the hose between each patient. If running through the autoclave, follow instructions listed above for Bite Block, Restorative Adapter & Short Hose.

Saqlash
Iltimos, Releaf® xona haroratida saqlanganligiga ishonch hosil qiling.
The Long Hose and Short Hose are designed to be held in the HVE Adapter when unused. (See image)

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (4)

VALIDATED STERILIZATION CYCLES

Gravitatsiyaviy joy almashish avtoklav

  • HARORAT: 121°C/250°F
  • MINIMAL EHM VAQTI: 30 daqiqa
  • MINIMAL QURUSH VAQTI: 30 min
VALIDATED STERILIZATION CYCLES
Gravitatsiyaviy joy almashish avtoklav Oldindan vakuumli avtoklav
TEMPERATURE: 121°C/250°F TEMPERATURE: 132°C/270°F
MINIMAL EHM VAQTI: 30 daqiqa MINIMAL EHM VAQTI: 4 daqiqa
MINIMAL QURUSH VAQTI: 30 min MINIMAL QURUSH VAQTI: 20 daqiqa

MUAMMOLARNI BARTARAF QILISH; NOSOZLIKLARNI TUZATISH

Proper Releaf® Restorative Position
Ensure the long end of the Restorative Leaf wraps around the retromolar area on the quadrant you are working on. To assist in evacuation, when a small amount of water and saliva build up behind the Restorative Leaf, have the patient slowly push their tongue against the Leaf while closing down on the Bite Block.
This will evacuate any remaining water that flows behind the Leaf.

Bo'shashgan aloqa
If you find the coupler becoming loose, it may be necessary to replace the O-ring.

Releaf® Innovative Dental Technologies, Inc.ning ro'yxatdan o'tgan savdo belgisidir.
Faqat stomatolog tomonidan yoki uning buyrug'i bilan sotilishi kerak.
Ushbu qurilmalardan ishlab chiqaruvchi tomonidan tavsiya etilgandan boshqa maqsadlarda foydalanish taqiqlanadi.

tomonidan ishlab chiqarilgan

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (5)INNOVATIVE PRODUCT BRANDS™
Innovative Product Brands Inc. TM

IPB, Inc

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (6)

TSS

Are there any contraindications for using the ReLeaf Restorative system?

No known contraindications have been identified.

What precautions should be taken while using the product?

Risk of allergic reactions cannot be excluded for those predisposed to allergies.

How should loose connections be addressed?

If you find the coupler becoming loose, it may be necessary to replace the O-ring.

Hujjatlar / manbalar

ReLeaf RESTORATIVE Basic Setup Kit [pdf] Qo'llanma
RESTORATIVE Basic Setup Kit, RESTORATIVE, Basic Setup Kit, Setup Kit, Kit

Ma'lumotnomalar

Fikr qoldiring

Sizning elektron pochta manzilingiz nashr etilmaydi. Majburiy maydonlar belgilangan *